![background image](http://html1.mh-extra.com/html/aandd/ad-4212a/ad-4212a_instruction-manual_446302021.webp)
19
Unit
AD-4212B-102
AD-4212B-201
AD-4212B-301
Capacity
Minimum
display
Capacity
Minimum
display
Capacity
Minimum
display
Gram
110
0.00001
210
0.0001
310
0.0001
Milligram
110000
0.01
210000
0.1
310000
0.1
Ounce (Avoir)
3.88
0.000001
7.40
0.00001
10.93
0.00001
Troy Ounce
3.53
0.000001
6.75
0.00001
9.96
0.00001
Metric Carat
550
0.0001
1050
0.001
1550
0.001
Momme
29.3
0.00001
56.0
0.0001
82.6
0.0001
Pennyweight
70.7
0.00001
135.0
0.0001
199.3
0.0001
Grain (UK)
1697
0.0002
3240
0.002
4784
0.002
Tael
(HK general, Singapore)
2.91
0.000001
5.55
0.00001
8.20
0.00001
Tael (HK jewelry)
2.93
0.000001
5.61
0.00001
8.28
0.00001
Tael (Taiwan)
2.93
0.000001
5.60
0.00001
8.26
0.00001
Tael (China)
3.52
0.000001
6.72
0.00001
9.92
0.00001
Tola (India)
9.43
0.000001
18.00
0.00001
26.57
0.00001
Messghal
23.4
0.00001
44.8
0.0001
66.1
0.0001
5-2 Changing the Unit
A unit or mode can be selected and stored in the function table.
Select a unit or mode as follows:
1 Press and hold the SAMPLE key until
ba5fnc
of the
function table is displayed, then release the key.
2 Press the SAMPLE key several times to display
Unit
.
3 Press the PRINT key to enter the unit selection mode.
4 Press the SAMPLE key to select a unit or mode to be
used for weighing.
5 Press the RE-ZERO key to specify the unit or mode
selected in step 4.
(The stabilization indicator
illuminates.)
6 Press the PRINT key to store the unit or mode. The
balance displays
end
and then displays the next
menu item of the function table.
7 Press the CAL key to exit the function table. Then the
balance returns to the weighing mode with the selected
unit or mode.
Select
Specify
Store
Displays the units
sequentially.