![A.M.I Italia Saver One P User Manual Download Page 68](http://html1.mh-extra.com/html/a-m-i-italia/saver-one-p/saver-one-p_user-manual_4342625068.webp)
68
13.10
P
ATIENT ANALYSIS SYSTEM IN SEMI
-
AUTOMATIC MODE
Category
Rated specifications
Function
It determines patient impedance and assesses the ECG rhythm and the quality of the signal, to
determine whether administering the shock is appropriate or not.
Impedance range
20 -
200 Ω
ECG analysis time
≥4 seconds (with new fully charged battery)
Sensitivity
97% Complies with IEC/EN60601-2-4 guidelines
Specificity
99% Complies with IEC/EN60601-2-4 guidelines
Shockable rhythms
If used on a patient with the characteristics listed in the use criteria, the
Saver One P
defibrillator is designed to recommend a defibrillating shock when it detects the right
impedance and when the following situations occur:
Ventricular Fibrillation
peak-to-peak amplitude at least 200µVolts
Ventricular tachycardia
with heartbeat frequency min. 180 bpm and peak-to-peak amplitude
at least 200µVolts (including ventricular flutters and polymorphic Ventricular tachycardia)
Non-shockable rhythms
The
Saver One P
is designed to not recommend shocks with all other rhythms, including:
normal sinusoidal rhythm, moderate ventricular fibrillation (<200 μVolts), some slow
ventricular tachycardias and asystoles.
13.11
ECG
A
NALYSIS OPERATION
ECG rhythm
Dimension
Test sample
Objective
Value detected
Rhythm to be defibrillated
Ventricular Fibrillation (VF)
500
Sensitivity > 90%
98%
Rhythm to be defibrillated
Ventricular Tachycardia (VT, bpm >140)
600
Sensitivity > 75%
92%
Rhythm not to be defibrillated
Normal sinus rhythm
1500
Specificity > 99%
100%
Rhythm not to be defibrillated
Asystole
30
Specificity > 95%
100%
Non-treatable rhythm
Generic AF, SVT, PVC
30
Specificity > 95%
100%
Positive predictive values
97.1%
False positives
4.1%
13.12
ECG
M
ONITORING
Category
Rated specifications
Type of protection
BF
with defibrillation PADs
CF
only with SAV-C0017 cable
Intended use
The ECG Monitoring function is intended to be used in situations where the patient is
always followed by an operator and is never left alone
Bandwidth
0.5 at 40Hz (default) with defibrillation PADs or ECG cable SAV-C0017
Lead ECG
Lead II
ECG Amplitude Range
10 mm/mV
Heart Rate Range
30
–
200 bpm
Heart Rate Resolution
1 bpm
Heart Rate Alarm (*)
Threshold MIN = 30 bpm; Threshold MAX 120 bpm
(*)
The alarm thresholds are set by the Manufacturer and cannot be changed locally. On request it is possible to have different
threshold values