5.3.4
Universal Face-to-Face SAV-C0599 Defibrillation PADs Management
The SAV-C0599 defibrillation PADs are managed by the models that provide for patient selection (see paragraph 6.1)
with the aid of the display: after the device is switched on, the display immediately shows the screen in which the
operator is asked to select adult or paediatric patient, using the menu navigation keys, which are designated
separately for the two types of patients for greater safety. Once the desired choice has been made, press the button
to confirm the selection.
• ARROW UP
ADULT
•
ARROW DOWN
CHILD
With the SAV-C0599 Universal PADs, the following discharges can be delivered after patient selection:
•
Adults: energy up to 200J (SVP-B0006-U, SVP-B0006-U-Q), up to 360J (SVP-B0007-U, SVP-B0007-U-Q)
•
Children:
maximum energy of 50J (on all compatible models see section 5.3).
5.3.5
Placement of the defibrillation pads
The correct placement of the PADs is essential for an efficient analysis of the patient's heart rhythm and for the
consequent delivery of the shock (if necessary).
For the placement and polarity of the electrodes of each type of pad, always refer to the instructions on the packaging
and in the instruction manual of the PADs.
▪
Emergency defibrillation
▪
Cardioversion
▪
Stimulation
▪
Monitoring (provides a Lead II
track)
▪
Emergency defibrillation
▪
Cardioversion
▪
Stimulation
▪
Monitoring