![A.M.I Italia Saver One P User Manual Download Page 42](http://html1.mh-extra.com/html/a-m-i-italia/saver-one-p/saver-one-p_user-manual_4342625042.webp)
42
8.5
S
HOCKABLE RHYTHM
If the
Saver One P
after analysing the patient's heart rhythm recognises a VF or VT. This information is highlighted by
voice (audio) and visual (colour display) prompts, as shown in the following table:
Voice prompts
Text
Video Display
Shock Recommended
Shock recommended
Then it automatically performs the charging phase. This information is highlighted by voice (audio) and visual (colour
display) prompts, as shown in the following table:
Voice prompts
Text
Video Display
Do not touch the patient
Charging
… in progress ..
Charging
The progress of the device charge is indicated by the charging bar
Once the charging phase is completed, the
Saver One P
is ready to shock. This information is highlighted by voice
(audio) and visual (colour display) prompts, shown in the table; in addition, the shock button light will blink
Voice prompts
Text
Video Display
Stay clear from patient
Press shock button
Push the blinking red button
Before pressing the shock button, move away and make sure that no one is touching the patient.
To shock, press the shock button within 15 seconds
If the shock button is not pressed within 15 seconds of the shock alert, the
Saver One P
will automatically disarm. This
information is highlighted by voice (audio) and visual (colour display) prompts, as shown in the following table:
Voice prompts
Text
The shock button was not pushed
Shock button not pressed
Shock Cancelled
Shock Cancelled