ROUBLESHOOTING GUIDE
The following table lists the symptoms, the possible causes and the possible corrective actions of the problems that
may arise. For more information on the implementation of corrective actions, refer to the other sections of the
operator's manual. If the failure of the unit persists, request assistance.
Symptom
LED
Mini DISPLAY
Colour TFT
Possible cause
Corrective action
Device with battery installed
does not switch on, the LED and
the control display are both off
The battery is completely
empty or faulty
Replace the battery. If the problem
persists, contact technical support
The device is not working
Please contact technical support
In
standby
the control LED
flashes green but the mini
display is off
The mini display is broken
Contact a support centre
In
standby
the control LED is off
but a "V" appears on the control
mini-display.
The control LED is broken
Contact a support centre
In
standby
the control LED
flashes RED and a wrench
appears on the control display
A critical error of the device
was found during the daily
self-test.
Contact a support centre and report the
error code.
In
standby
the control LED
flashes GREEN/RED alternately
and a wrench appears on the
control display
Battery empty
Level <1%
The device could switch off
during use.
(see the relevant paragraph)
Replace the battery
In the
operating mode
the voice
prompt
"Low
battery"
is
provided
Battery low.
Battery level 5%.
It is possible to use the device
but the battery level is low
(see the relevant paragraph)
Purchase a new battery and replace it as
soon as possible.
During normal use the voice
prompt "Battery low, Replace" is
issued
The battery is empty.
Level <1%
The device could switch off
during use.
(see the relevant paragraph)
Avoid using the device if possible.
Replace the battery
With the device turned on and
after placing the PADs on the
patient, the device continues to
prompt:
“Place the Pads”
The PADs connector has not
been inserted correctly or it
has been removed
Insert the PADs connector in the
appropriate compartment
The Pads have been placed
incorrectly
Please place the Pads correctly on the
patient’s naked chest. If needed, remove
chest hair with a razor
The Pads are faulty
Please control that the Pads are intact and
their expiry date; replace them, if needed
When inserting the battery the
Activation test
requires you to
press the shock button to start
the test. The button is pressed
but the test does not start. For
about 60 seconds the AED
requires to press the button and
then it turns off automatically,
signalling “Error xx” on the mini
LCD.
The shock button does not
work properly
Try switching off the device and repeating
the test. If the problem persists, contact
technical support
The device turns on, the mini
display and the TFT are on but no
voice prompt is issued
The device’s speaker is not
working
Please contact technical support
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF