TMA Grand Master 2.6 – User’s Guide
4 Operating the Main Unit
Important!
Before starting a larger project it is advised to calculate the number of cores that will
be inserted to PCR tubes, so placing a sufficient number of cleaning blocks
accordingly.
The number of cleaning spots depending on the puncher diameter are the following:
Puncher diameter
Number of cleaning spots
0.6mm
252
1mm
144
1.5mm
72
2mm
50
4.4.3 Inserting a block into the main unit
1. Recipient block positions are displayed in a double-column on the left side of the screen. Tray
positions of donor blocks are identified by six buttons (
A-F
), tray capacity is 10 blocks each.
December 9, 2016 – Rev. 1
3DHISTECH Ltd.
30(92)
2
1