45
CG000542 | Rev A
User Guide | CytAssist Instrument | Maintenance
If the Visium CytAssist will be transported in a vehicle or shipped,
it should be repackaged according to the instructions below. Run a
new Readiness Test (see
section) after relocating the
instrument.
In the event of an instrument return to 10x Genomics, Support will
guide the user through the correct documentation, in addition to the
packaging procedure below.
Step 1: Shipping Mode
a.
Launch Shipping Mode (for safe instrument transportation).
Menu Options > System > Alternate Modes > Shipping Mode
b.
Once Shipping Mode is active, follow onscreen instruction to power
down the instrument.
c.
Turn off the instrument completely using the power switch on the
back of the instrument.
d.
Remove the power cord.
e.
Decontaminate the instrument (see
section).
Step 2: Packaging Instructions
a.
Fully tighten the two adjustable feet at the front of the instrument by
turning them clockwise.
b.
Retrieve the Shipping Bolt from storage. Insert and tighten the
Shipping Bolt at the back of the instrument (the Shipping Bolt holds
the camera focus in a safe position during transportation).
Moving & Shipping
the Instrument
Shipping
Bolt
Adjustable Feet
Summary of Contents for Visium CytAssist
Page 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications...
Page 8: ...Safety Compliance Information 9 Safety 10 Regulatory...
Page 11: ...System Components 12 System Components 13 Accessory Kits...
Page 24: ...Menu Options 23 Menu Options 26 Firmware Update Procedure...
Page 42: ...Maintenance 41 System Maintenance 43 Moving Shipping the Instrument...
Page 48: ...Troubleshooting 47 Troubleshooting 49 Error Screens...