50
CG000542 | Rev A
User Guide | CytAssist Instrument | Troubleshooting
•
Instrument is hot
: If the Visium CytAssist overheats during an
experiment run, the run will fail. A warning message will appear
instruting users to carefully remove slides and contact
•
Poor tissue alignment on Visium CytAssist Slide
: Refer to the
section to verify tissue placement on a tissue slide.
Practice using the Visium CytAssist Training Kit and User Guide
(CG000549).
• Readiness Test Fails:
Refer to the
section and clean
the instrument as described. Wipe down the Tissue Slide Stage and
the Visium Slide Stage with 5-10 % bleach solution followed by 70%
ethanol or 70% isopropanol. Use compressed air to dry and remove
debris as needed. Clean glass windows with dry polyester swabs (such
as Texwipe TX714A Cleanroom Swab, 0.5”L - EW-33677-62 from Cole-
Parmer or equivalent) and compressed air. Use a bright light (e.g.
flashlight) to inspect the glass windows for any smudges and remove
them as described above.
If the problem persists, contact
Tissue Placement Relative to 6.5 mm Alignment Guides
Troubleshooting
contd.