35
CG000542 | Rev A
User Guide | CytAssist Instrument | System Operation
f.
Proceed to Leveling Mode. To level the instrument, turn the adjustable
feet at the front of the instrument. Adjust left-right tilt first by adjusting
one foot first and adjust the front back tilt using both feet at the same
time until the both bars turn green.
Turn feet clockwise for decreasing and counterclockwise for increasing height. The
feet will come off the instrument if they are continued to be turned counterclockwise.
Once the instrument is leveled correctly, check the back of the instrument to confirm
that the center back non-adjustable foot is resting on the table top while the two
non-adjustable corner back bumpers are not touching the table top.
See
section if an error occurs during the first-time setup protocol.
Back of a Leveled Instrument
Center foot touching table top
(non-adjustable)
Corner back bumper
not touching table top
Corner back bumper
not touching table top
Step 2:
First-Time Use
contd.