40
CG000542 | Rev A
User Guide | CytAssist Instrument | System Operation
g.
Press the play button to start the run.
• Midrun progress bar will show the time remaining in the run.
• Yellow before a run begins indicates a user-recoverable error (see
).
h.
At the end of a run, the button will display “Done” and a “Run Info”
tab at the bottom of the screen.
• Green indicates a successfully completed run.
• Red indicates a failed run/error (see
).
• Yellow indicates that the run has been canceled.
i.
Click the “Done” button and open lid. DO NOT power off the
instrument at this time, as it needs to process support data.
j. IMMEDIATELY
remove the Visium CytAssist slide from the instrument,
and proceed with the protocol workflow as specified in the
appropriate assay kit user guide (save cleanup for later).
Step 4:
Main Operation
contd.