7
CG000542 | Rev A
User Guide | CytAssist Instrument | Introduction
Product
Identification Label
Parameter
Specifications
Weight
~18.3 lb (8.3 kg)
Instrument Dimensions with:
Lid closed
Lid open
Maximum height
W D H
8”[203 mm] × 12.3”[313 mm] × 9.76”[248 mm]
8”[203 mm] × 12.3”[313 mm] × 12.3”[312 mm]
13.1” [334 mm]
Electrical Requirements
Nominal
from a standard 3-prong wall receptacle
that includes a safety ground pin
100-240 VAC, 50/60 Hz, 250 W
80-264 V operational range (+/- 10% of nominal)
Over voltage Category II (standard receptacle)
Pollution
Degree 2 (Indoor Use Only)
Ventilation Requirement
Minimum 4” [10 cm]
Around all sides
Operating Temperature
64-82°F [18-28°C]
Use in a typical indoor laboratory environment.
Extreme temperature conditions will affect the
sensitive reagents used with the instrument.
Humidity
30-80% R.H. non-condensing
Power Cable Length
Power cables will be in accordance
with regional specifications
6-7 ft [2 m] Standard
Product
Specifications
9.76”
8”
12.3”
13.1” max height
21
© 10X GENOMICS, INC. 2020 | CONFIDENTIAL – DO NOT DISTRIBUTE.
Visium CytAssist LBL-00042-01 Regulatory Label
● Must be legible from 18
inches
● Electrical shock warning not
required, but included. Must
be in English and French if
certified for Canada
●
Instrument SN: CA is an approved
product code, starting SN will be
CAVG10001
Image of label for reference only
Instrument Serial Number
21
© 10X GENOMICS, INC. 2020 | CONFIDENTIAL – DO NOT DISTRIBUTE.
Visium CytAssist LBL-00042-01 Regulatory Label
● Must be legible from 18
inches
● Electrical shock warning not
required, but included. Must
be in English and French if
certified for Canada
●
Instrument SN: CA is an approved
product code, starting SN will be
CAVG10001
The instrument requires a
~20 min cool down
between runs.
Summary of Contents for 1000433
Page 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications ...
Page 8: ...Safety Compliance Information 9 Safety 10 Regulatory ...
Page 11: ...System Components 12 System Components 13 Accessory Kits ...
Page 24: ...Menu Options 23 Menu Options 26 Firmware Update Procedure ...
Page 42: ...Maintenance 41 System Maintenance 43 Moving Shipping the Instrument ...
Page 48: ...Troubleshooting 47 Troubleshooting 49 Error Screens ...