EID for Non-Life-Support Functions
9650-000272-05 Rev. B
SurePower Single Bay Charger Operator’s Guide
A-3
EID for Non-Life-Support Functions
The non-life-support functions of the
SurePower Single Bay Charger
are intended for use in the
electromagnetic environment specified below. The customer or the user of the
SurePower
Single Bay Charger
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
20 V/m
80 MHz to 2.5 GHz
3 Vrms
20 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
SurePower Single Bay Charger
,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d = 1.2
150 KHz to 80 MHz
d = 0.35
80 MHz to 800 MHz
d = 0.7
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer, and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
NOTE 1 At 80 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the
SurePower Single Bay Charger
unit
is used exceeds the applicable RF compliance level above, the
SurePower Single Bay
Charger
unit should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
SurePower Single-Bay Charger unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P
Содержание SurePower Single Bay Charger
Страница 1: ...SurePower Single Bay Charger Operator s Guide 9650 000272 05 Rev B...
Страница 14: ...This page intentionally left blank...
Страница 20: ...This page intentionally left blank...
Страница 34: ...This page intentionally left blank...