Testing Batteries (Manual Procedure)

The charger’s battery test recalibrates the battery and measures the battery’s charge capacity to 
ensure that the battery can hold a charge that is sufficient for effective service, which is at least 
60% of its original factory charge capacity.

When should you Test and Recalibrate a Battery?

The 

SurePower and SurePower II Battery Packs 

automatically recalibrate and measure charge 

capacity after a set service interval (1 year or 500 Amphere-hours discharged)

You can also 

initiate a manual battery test, which recalibrates the battery and measures charge capacity, after 
a shorter service interval (for example, after a specific number of charge/discharge cycles or 
less than a year after the last recalibration). 

Running a manual battery test resets the interval after which the battery 
automatically recalibrates and measures charge capacity.

Note:

A battery will pass a battery test (the charger lights the 

Ready

 LED) if the battery can 

be charged to at least 60% of its original factory charge capacity. If manual test results 
in a charging fault, you should remove the battery from service and perform additional 
testing (see the 

Troubleshooting

 section in Chapter 4.)

Содержание SurePower Single Bay Charger

Страница 1: ...SurePower Single Bay Charger Operator s Guide 9650 000272 05 Rev B...

Страница 2: ...information updates are available SurePower and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the prope...

Страница 3: ...ePower II 1 2 Battery Bay Panel 1 3 Charger and Battery Keys 1 4 Battery Charging Requirement 1 4 Charging Modes 1 5 QuickCharge 1 5 ManualTest 1 5 Chapter 2 Set Up Powering on the Charger 2 2 Self Te...

Страница 4: ...nd SurePower II Battery Pack 5 1 Developing a Battery Management Program 5 1 Chapter 6 Product Specifications SurePower Single Bay Charger Appendix A Manufacturer s Declaration Electromagnetic Emissio...

Страница 5: ...s SurePower Single Bay Charger which you use to test recalibrate and charge ZOLL s rechargeable lithium ion defibrillator batteries It is important that you read and understand the information in thi...

Страница 6: ...r 3 Operation Describes how to use the SurePower Single Bay Charger to test recalibrate and charge ZOLL s rechargeable defibrillator batteries Chapter 4 Maintenance and Troubleshooting Describes how t...

Страница 7: ...er to Operator s Guide for important safety related information such as Warnings and Cautions that cannot be presented on the device DANGER high voltage present Protective earth ground terminal Altern...

Страница 8: ...face type Return to a collection site intended for waste electrical and electronic equipment WEEE Do not dispose of in unsorted trash With respect to electric shock fire mechanical and other specified...

Страница 9: ...container is damaged also notify the carrier Contacting Technical Service The SurePower Single Bay Charger does not require periodic recalibration or adjustment You should however periodically perform...

Страница 10: ...the SurePower Battery Pack is AG The first two characters of the date of manufacture code give the last two digits of the year for example 05 would appear for products manufactured in 2005 The last c...

Страница 11: ...Use only the AC cord supplied with the device Failure to use the proper line cord could result in excess leakage currents EMC problems and reduced safety Do not disassemble the SurePower Single Bay C...

Страница 12: ...Monitor Defibrillators or a SurePower Single Bay Charger that is not plugged into live AC mains when storage may exceed 90 days Battery damage may occur Do not connect any cables or devices to the co...

Страница 13: ...lmsford MA 01824 4105 Fax 978 421 0025 Tel 978 421 9655 Notification of Adverse Events As a health care provider you may have responsibilities under the Safe Medical Devices Act SMDA for reporting the...

Страница 14: ...This page intentionally left blank...

Страница 15: ...ingle Bay Charger is a single bay unit that can test recalibrate and charge a single ZOLL rechargeable defibrillator battery Figure 1 1 SurePower Single Bay Charger The charging bay in the SurePower S...

Страница 16: ...arger with SurePower Battery Pack The following illustration is of a SurePower Single Bay Charger with a SurePower Battery Pack installed in it Charger with SurePower II Battery Pack The following ill...

Страница 17: ...to indicate that the battery is charging Ready The LED lights next to this icon to indicate that the battery is fully charged and ready to use Test button Pressing this button initiates a manual test...

Страница 18: ...ion shows the key on the SurePower Single Bay Charger Figure 1 3 Battery Key Batteries that do not have the correct keys will not seat in the charging bay and cannot be charged and tested by the SureP...

Страница 19: ...harging method for ZOLL s lithium ion batteries such as the SurePower Battery Pack The QuickCharge mode is the quickest way to charge lithium ion batteries which do not require frequent testing to mai...

Страница 20: ...This page intentionally left blank...

Страница 21: ...ion Figure 2 1 SurePower Single Bay Charger Setup Caution Do not place anything on top of or beneath the unit that might block the vents Do not remove or discard the charger s port connector plug remo...

Страница 22: ...power is available Position the SurePower Single Bay Charger so that the AC cord is easy to disconnect Warning Always inspect the AC cord before use Do not use the AC cord if it is defective for examp...

Страница 23: ...SurePower Single Bay Charger Operator s Guide 3 1 Chapter 3 Operation The SurePower Single Bay Charger is designed to charge and recalibrate two different types of batteries the SurePower and SurePow...

Страница 24: ...ce the back end of the battery pack against the back of the charging bay where the pins are not located and then press the battery down firmly until it clicks Warning To avoid pinching your fingers ho...

Страница 25: ...te after you insert the battery 3 Allow the battery time to charge The charging time will depend on the battery type how depleted the battery was when you put it into the charger and which charging me...

Страница 26: ...lip to disengage it from the latch and then lift the battery out of the charging bay Note If the Fault LED is on when you remove the battery pack the charger beeps to alert you that there has detected...

Страница 27: ...clean undamaged charger 3 If the battery pack and charger are clean and undamaged reinsert the battery pack 4 If the Fault LED lights after reinserting the battery pack remove the battery pack wait 1...

Страница 28: ...Amphere hours discharged You can also initiate a manual battery test which recalibrates the battery and measures charge capacity after a shorter service interval for example after a specific number of...

Страница 29: ...the bay s control panel The charger lights the Test LED and begins testing the battery The Test LED will remain lit for the duration of the test and the Charging LED during the charging portion of th...

Страница 30: ...anel s LEDs can be tested to determine if they are working correctly To test the control panel s LEDs follow this procedure Step Action 1 If there is a battery pack in the charging bay remove it 2 Pre...

Страница 31: ...what the LED and alarm activity indicates during different operating modes The charger light LEDs in one of two ways Steady persistent or brief periodic Indicators When Powered On or Testing the Unit...

Страница 32: ...nual Test Method battery charging On steady On steady Battery Capacity Test successful Manual Test Method On steady On steady Battery Capacity Test successful Automatic Recalibration On steady On stea...

Страница 33: ...lators and monitors Disabling Enabling the Audible Alarm You can disable and enable the charger s audible alarm To disable or enable the audible alarm press and hold down the charger s Test button for...

Страница 34: ...This page intentionally left blank...

Страница 35: ...eep turn on the fan and light all LEDs for several seconds The Power LED will remain lit after the other LEDs go off 4 With the battery bay empty check that all LEDs on the control panel except the Po...

Страница 36: ...ed in the bay Check that the contacts are clean and not damaged Use another battery that you know is good 3 Test LED does not light when you press the Test button Check that the battery is properly se...

Страница 37: ...ePower or SurePower II Battery Packs in a fully depleted state this will damage the battery Never store the SurePower or SurePower II Battery Packs in a charger that is not powered on Developing a Bat...

Страница 38: ...cardiopulmonary resuscitation should a battery pack or resuscitation device fail and need to be replaced 6 Test Battery Packs regularly Develop a testing schedule as part of your organization s Batte...

Страница 39: ...Current 0 to 2 5 A dc Maximum Thermal Output 70 BTU HR Design Standards Designed to meet or exceed IEC 60601 1 standards for medical equipment safety Temperature Operation 32 F to 122 F 0 C to 50 C S...

Страница 40: ...Maximum Charge Time SurePower and SurePower II Batteries 4 5 hours at 25 C 77 F 8 5 hours at 50 C 122 F Battery Contact Pins Usage Limit 5 000 removal insertion cycles...

Страница 41: ...such an environment RF emissions CISPR 11 Group 1 The SurePower Single Bay Charger does not contain any intentional RF transmitters Therefore its RF emissions are very low and are not likely to cause...

Страница 42: ...e that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip i...

Страница 43: ...ion distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference ma...

Страница 44: ...unications equipment Rated Maximum Output Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 0 35 800 MHz to 2 5 GHz d 0 7 0...

Отзывы: