9650-000272-05 Rev. B
SurePower Single Bay Charger Operator’s Guide
3-5
Troubleshooting Charging Faults
If the charger detects a charging fault, you can perform the following tasks to determine the
nature of the fault:
1.
Remove the battery pack and inspect it.
If the battery pack’s casing or pins are damaged,
remove that battery from service.
2.
Inspect the charging bay.
If the charging bay or the its contacts are dirty, unplug the charger
from the AC mains, clean it (following the procedure that we describe in Chapter 4), and then
reinsert the battery pack. If the charger casing or the charging bay is damaged, unplug the
charger from the AC mains, remove it from service, and then test the battery in a clean,
undamaged charger.
3.
If the battery pack and charger are clean and undamaged,
reinsert the battery pack
.
4.
If the
Fault
LED lights after reinserting the battery pack,
remove the battery pack, wait 10
seconds, and then press the battery pack’s Test button.
If the battery’s
Fault
LED lights,
remove the battery pack from service. If the battery pack’s fault light does not light, test the
battery pack in another charger.
Removing the SurePower II Battery Pack -- lift the latch on the SurePower II Battery
Pack. The battery pack will pop out of the charging bay.
Note: If the Fault LED is on when you remove the battery pack, the charger beeps to alert
you that there has detected a charging fault. Do not use this battery pack n a
defibrillator or monitor until you have confirmed the nature of the charging fault (for
more information, see Troubleshooting Charging Faults later in this chapter.