9650-000272-05 Rev. B
SurePower Single Bay Charger Operator’s Guide
3-7
Performing a Manual Battery Test
To perform a manual battery test, follow this procedure:
Step
Action
1.
Insert the battery pack into the charging bay and press the
Test
Button on the bay’s control panel
The charger lights the
Test
LED and begins testing the battery. The
Test
LED will remain lit for
the duration of the test
...and the
Charging
LED during the charging portion of the test:
When the battery is fully charged and has passed a factory-configured minimum capacity test,
the charger lights the
Ready
and
Test
LED’s
If the charger determines that the battery has insufficient capacity, it lights the
Fault
and
Test
LED’s