Responding to alarms
Page 5-15
Check belt message
After you get a number of belt problem screens, you may get this
message. It means that the monitor is still not receiving a good signal
from the electrode belt.
If you get this message:
1 Disconnect the electrode belt from the monitor.
2 Take off the garment and make sure that nothing is covering the
metal surfaces of the ECG electrodes.
3 Put a dab of unscented hand lotion, skin cream, or moisturizer on the
surface of each ECG electrode (the round ones, not on the therapy
pads).
4 Put the garment on and connect the electrode belt to the monitor.
While you’re wearing the LifeVest, tap the help button
for
reminders about what to do.
If problems continue, call ZOLL.
Содержание LifeVest 4000
Страница 1: ...LifeVest Model 4000 Patient Manual PN 20B0047 EUK Rev M May 1 2021...
Страница 6: ...LifeVest 4000 Patient Manual Page vi This page intentionally left blank...
Страница 60: ...LifeVest 4000 Patient Manual Page 3 36 This page intentionally left blank...
Страница 76: ...LifeVest 4000 Patient Manual This page intentionally left blank Page 4 16...
Страница 98: ...LifeVest 4000 Patient Manual Page 5 22 This page intentionally left blank...
Страница 104: ...LifeVest 4000 Patient Manual 7 When you get this message you have finished the health survey Tap OK Page 6 6...
Страница 124: ...LifeVest 4000 Patient Manual This page intentionally left blank Page 6 26...
Страница 127: ...Quick charts Airplane mode When you re finished with device If you get an alert Page A 3...
Страница 130: ...LifeVest 4000 Patient Manual This page intentionally left blank Page B 2...
Страница 138: ...LifeVest 4000 Patient Manual This page intentionally left blank Page D 2...