Using the LifeVest
Page 3-11
How to read the charger touchscreen
If your charger has a touchscreen, the screen shows the status of the battery and
charger. When the room gets dark, the charger display gets dim; see next page.
Screen shows
What it means
What to do
There is no battery in the
charger
Insert the battery into the charger. Leave one battery in the
charger while you use the other battery.
Battery is charging
Let the battery charge. This can take up to 4 hours.
Battery is fully charged,
ready for use in monitor
Leave battery in the charger until ready to exchange battery.
Battery is being tested
Let the test run, which can take up to 12 hours. Test verifies
battery function as part of normal routine. After test, battery
will be charged, which can take another 4 hours. So the whole
process can take up to 16 hours.
If you need to use the battery sooner, remove battery from
charger as soon as you see this on the screen. Then put the
battery back into the charger to skip the test and charge the
battery. The test will be performed the next time the battery is
plugged into the charger.
Battery has a problem, but
still might hold a charge
The battery may charge, but will take longer than normal.
Battery can be used, but should be replaced.
Call ZOLL for a replacement battery.
Charger has a problem, and
cannot charge the battery.
Do not leave a battery in the charger. Battery is not being
charged.
If the problem continues, call ZOLL.
Содержание LifeVest 4000
Страница 1: ...LifeVest Model 4000 Patient Manual PN 20B0047 EUK Rev M May 1 2021...
Страница 6: ...LifeVest 4000 Patient Manual Page vi This page intentionally left blank...
Страница 60: ...LifeVest 4000 Patient Manual Page 3 36 This page intentionally left blank...
Страница 76: ...LifeVest 4000 Patient Manual This page intentionally left blank Page 4 16...
Страница 98: ...LifeVest 4000 Patient Manual Page 5 22 This page intentionally left blank...
Страница 104: ...LifeVest 4000 Patient Manual 7 When you get this message you have finished the health survey Tap OK Page 6 6...
Страница 124: ...LifeVest 4000 Patient Manual This page intentionally left blank Page 6 26...
Страница 127: ...Quick charts Airplane mode When you re finished with device If you get an alert Page A 3...
Страница 130: ...LifeVest 4000 Patient Manual This page intentionally left blank Page B 2...
Страница 138: ...LifeVest 4000 Patient Manual This page intentionally left blank Page D 2...