9650-000752-01 Rev. D
ZOLL AED 3 Administrator’s Guide
25
Optional Maintenance Test for Technical
Professionals
Although the ZOLL AED 3 automatically performs maintenance testing during
periodic self tests, you can periodically perform the following manual test to
verify that the ZOLL AED 3 is working properly and is ready for use.
This section has two tests: one for ZOLL AED 3 and ZOLL AED 3
BLS
models,
and one for ZOLL AED 3
Automatic
models.
Test Procedure (ZOLL AED 3 and ZOLL AED 3
BLS
Models)
NOTE
You must use the ZOLL AED Simulator with this test.
1.
Connect the ZOLL AED Simulator to the ZOLL AED 3 defibrillation pads con-
nector.
2.
Power on the simulator and the ZOLL AED 3 defibrillator. Verify that all of
the following occur:
•
The status indicator (located on the top left of the AED) initially changes
from blank to a green check (
)
within 4 to 5 seconds after the AED is
turned on.
•
The AED issues the
UNIT OK
voice prompt and displays the message within
5 seconds after power-up.
•
The AED displays the number of shocks in the bottom left of the screen and
the event elapsed time (since power-up) in the bottom right of the LCD
screen.
3.
Using the simulator, input a VF rhythm to the AED. Verify that after the AED
proceeds through its sequence of victim assessment prompts, it does the
following:
•
Analyzes the ECG rhythm.
•
Issues the
SHOCK ADVISED
voice prompt.
•
Charges the defibrillator.
•
Issues the
DON’T TOUCH THE PATIENT; PRESS THE FLASHING
SHOCK
BUTTON
voice prompt.
4.
Verify that the AED emits the charge ready tone and that the Shock button
illuminates.
5.
Press the Shock button and verify that the simulator shows that a shock
was delivered. Verify that the number 1 displays next to the shock icon on
the bottom left of the LCD screen.
6.
Following shock delivery, verify that the AED issues the
START CPR
mes-
sages.
7.
Activate the simulator’s CPR function. Verify that the metronome begins to
beep and that the following voice prompts/messages are issued within 60
seconds:
PUSH HARDER
followed by
GOOD COMPRESSIONS.
Содержание AED 3
Страница 1: ...Administrator s Guide 9650 000752 01 Rev D...
Страница 20: ...12 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 36: ...28 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 40: ...32 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 72: ...64 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...