9650-000752-01 Rev. D
ZOLL AED 3 Administrator’s Guide
37
The ZOLL AED 3 defibrillator is intended for use in the electromagnetic
environment specified below. The customer or the user of the AED should assure
that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
± 1 kV differential
mode
+/- 2 kV common
mode
Not applicable
Not applicable
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
<5%
U
T
(>95%
dip in
U
T
) for 0.5
cycle
40%
U
T
(60% dip
in
U
T
) for 5 cycles
70%
U
T
(30% dip
in
U
T
) for 25
cycles
<5%
U
T
(>95%dip in
U
T
)
for 5 sec
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the AED, including cables,
than the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter, or 30
cm, whichever is greater.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz outside ISM
bands
a
3 Vrms
d = 1.17
10 Vrms
150 kHz to
80 MHz in ISM
bands
a
10 Vrms
d = 1.20
P
P
Содержание AED 3
Страница 1: ...Administrator s Guide 9650 000752 01 Rev D...
Страница 20: ...12 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 36: ...28 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 40: ...32 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...
Страница 72: ...64 ZOLL AED 3 Administrator s Guide 9650 000752 01 Rev D...