27
©2019 Guangzhou ZHIYUAN Electronics
Date: 2019/01/17
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
ZDS2000B Series Oscilloscope User Manual
Scale
14div
(
horizontal
)
× 8div
(
vertical
)
4.10 Input/output Port
Table 4-12: Input/output Port
Port Type
Description
USB HOST
Connect U disk
USB DEVICE
Connect PC
LAN
RJ-45 connector
,
support 10/100BASE-T
RS232
RS232
communication port
Trig Out
The BNC connector on the rear panel provides pulse output when the
oscilloscope is triggered.
Probe compensation output
Front panel pins; amplitude: 3.0V, frequency: 1KHz
4.11 Technical Specifications
Table 4-13: Technical Specifications
Power Supply
Description
Supply voltage
100
—
240V~
Supply frequency
50
—
60 Hz
Power
100W max
Fuse
5A, Class T, 250V
Mechanical Specifications
Description
Dimensions
Width*height*depth =
434mm × 221mm × 134mm
Weight
Net weight: 4Kg; gross weight: 6.5Kg
Environment
Description
Temperature range
Operation
:
10
℃
~+40
℃
, Storage
:
-20
℃
~+70
℃
Ingress Protection Rating
IP20
Humidity range
≤60
%
relative humidity
Cooling method
Fan
Altitude
Less than 3000m (operating), less than 12000m (non-operating)
Electromagnetic Compatibility
2014/30/EU, EN61326-1:2013
Safety
Notes1, Notes2
IEC/EN61010-1: 2010,61010-1: 2010, IEC/EN 61010-2-030:2010
GB4793.1-2007
Pollution degree
2
Notes 1: This product is only used in non-residential areas. Use in residential areas may cause electromagnetic inte
rference.
Notes 2: In situations where the external electromagnetic field interference is serious (such as the radiated and con
ducted interference in the IEC61000-4-3 / 6 standard), if the user uses the alligator clip to ground to measure the si
gnal, the oscilloscope probe will easily couple to the external electromagnetic field interference, resulting in The os
cilloscope shows increased peak-to-peak noise and changes in horizontal waveform offset. At this time, the user sh
ould use the oscilloscope ground spring method for measurement to reduce the coupling of the probe measurement