Error Messages
Troubleshooting
Disposal
15
Page 23
Device Malfunction
No response to the main switch / display remains dark.
Possible cause 1
Mains power connection
Fix for cause 1
Check whether the power cord is properly plugged into the electrical outlet and
whether the plug is firmly inserted into the socket of the device.
Check the power cord for damage.
Check the power supply and mains socket.
Possible cause 2
Fuse
Fix for cause 2
In the power input socket of the device, there are micro-fuses that disconnect the
power supply in the event of an electrical problem. Open the cover and check the
fuses. If necessary, replace the faulty fuse.
Replace the fuse only with fuses of exact the same name or equivalent.
First, check the power supply completely for possible errors.
If the fault persists, contact the service representative / customer service
immediately
Error message SD
card
On the SD-card, the user-defined settings as well as the application list are stored.
If the SD-card is not inserted, the display application recommendation will not open
upon activation of the start button on the home screen.
The following message appears:
“SD card not found. Use of treatment recommendations requires a SD-card.”
Insert the card and restart the device.
If you lose your card, please contact the customer service department.
Water tank empty
The message below shows the user if there is not enough water in the tank
Please start to fill the tank according to chapter 14., if error persists
,
please contact
the customer service department.
The error message below shows the user that the filling process was stopped and
has to be repeated.