Basic Settings
7
Page 9
Protocol
Screen
(1) Energy
Displays the set pulse energy. When the application is active, the bar graph is filled in.
The pulse energy can be set before and during the delivery of pulses. The pulse
energy is adjustable between 60–185 mJ in 10 mJ steps.
(2) Frequency
(3) Handpiece
Displays the set frequency.
Frequency range: 1–22 Hz, adjustable via the right Encoder. The maximum selectable
frequency depends on the energy level selected.
Connection of handpiece.
(4) Mode
Displays the selected operating mode. When activating the “mode” window, the
selection menu appears with the operating modes:
Series pulses, burst 4 pulses, burst 8 pulses, burst 12 pulses.
The desired mode of operation is selected directly in the corresponding row.
(5) Number of
pulses
Display of the preselected number of pulses and the currently emitted pulses as well
as the total number of emitted pulses if the total number of emitted pulses is not pre-
selected. Display of the counting direction in ascending or descending order.
Activation of the “preselection” field will open the input menu for entering a
preselected number of pulses. The pre-selection can be done in 100 or 1000 steps.
(6) Counting
direction
Activation of the button determines the counting direction (ascending or descending)
of the number of pulses delivered.
(7) Reset
(8) Save
Resets to 0 in the case of ascending counting order and to the pre-set number of
pulses in the case of descending counting order.
Activation of the button opens the field for entering an individual program for storage
in the favorites list.
(9) Title bar
Display of the current program name and the selected handpiece.
9
2
3
4
5
6
7
8
1