Cleaning / Disinfection
10
Page 14
Silicone protection cap
Cleaning, manually:
Tools:
Drinking water, lukewarm
Vessel, e.g. kidney dish
Brush, e.g. medium hard toothbrush
Alcohol-free plastic cleaner (e.g. cleaner for medical devices)
Remove the silicone protection cap from the applicator head prior to cleaning. Prepare
a solution of the cleaning agent following the instructions of the manufacturer. Put the
silicone protection cap in the solution. Use the brush to clean all the inner and outer
surfaces of the protection cap. Finally rinse the protection cap under running water.
Disinfection, manual:
Tools:
Vessel, e.g. kidney dish
Commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, virucidal and fungicidal properties.
We recommend that disinfection is carried out at least once a week, as well as if there
is any indication of contamination. Consult with your hygiene specialist when doing so.
Always perform cleaning prior to disinfection.
Prepare a solution of the disinfection agent following the instructions of the
manufacturer. Put the silicone protection cap in the solution. Make sure that all the
inner and outer surfaces of the protection cap are coated. Leave the protection cap in
the solution for the time specified by the manufacturer of the disinfection agent. Finally
rinse the protection cap under running water.
.