Function Test
12
Page 25
After the device is switched on, the Z
Wave
Pro
performs a self-test, which checks all
internal components.
In the event of a fault, an error message appears.
In addition, an extended function test can be performed as described below.
This test should be performed monthly or if there are any doubts about the viability of
the device.
Note:
Before performing the functional test, ensure that the hand piece and foot switch are
connected properly to the device.
Check to see whether the power cable is connected properly to the device and
whether it is connected to the network.
Function test
Switch the device on.
Briefly press the foot switch – the fan and pulse generator start immediately. The pulse
generator must work at the frequency on the display (10 Hz as standard value).
Note:
Upon completion of the function test,
switch off the Z
Wave
Pro
.
If an application is to be carried out directly afterwards, set the desired application
parameters and proceed further as described in section 7.