Application Information
2
Page 2
Prior to using of the device on the patient, the user should become acquainted with the
operation instructions and individual treatment methods as well as the indications /
contraindications, warnings and other instructions. Additional sources of information
about types of therapy should be consulted.
Caution
!
Before use, ensure that the device is operated using a properly earthed socket with
earthing contact (electrical installation according to DIN VDE 0100 Part 710 or
equivalent). The device must be only operated with the supplied power cord. The
power cord must be protected against mechanical stress.
Caution
!
Operating the device near strong electromagnetic fields (e.g. tomography, x-ray, or
diathermy equipment) may interfere with the operation of the device. Please keep a
safe distance of several meters.
Z
Wave
Pro
is not suitable for use in explosive, flammable, or combustive environments.
During use, the device must be located in a position that allows direct access to the
mains power supply so that this can be disconnected at any time.
To prevent the risk of electric shock, the plug must be disconnected from the power
supply before performing any maintenance or cleaning activities.
Inspect the device before use. If there is any damage, do not use the device.
Caution
!
Only use accessories from Zimmer MedizinSysteme GmbH.
Caution
!
The handpiece of the device is not designed for continuous operation. After a max.
6000 strokes, a treatment pause of 15 min is required.
Caution
!
If Z
Wave
Pro
is not mounted on a designated wagon, ensure that the Z
Wave
Pro
is
placed on a stable surface.
Caution
!
To avoid heat accumulation in the handpiece, ensure that the vents on the top and
especially the bottom of the handpiece are not blocked (i.e. by the hand or otherwise).