Safety and Maintenance
11.2 Maintenance
11
Page 24
Before cleaning and maintenance, the device must always be switched off with the
main switch, and the power cord must be disconnected.
Checking the impact
dome
At the applicator head, a deformation / shortening of the rear dome takes place during
use. On a small scale, this will have no effect on functionality.
In cases of more extensive deformation of the dome or considerable shortening, the
applicator head must be replaced.
To check the condition of the applicator head, a test template is supplied with the
device, which can be used to check whether the wear limit has been reached.
Depending on the frequency of use, the device should be inspected at least once per
month.
To do so, the applicator head is removed and measured with the enclosed test
template.
Air gap
Applicator OK
Test template
Test template
Template is lying or air gap at the top
Wear limit reached