Manufacturer's EMC Declaration
18
Page 35
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device
ZWave
Med
is intended for use in the electromagnetic environment specified below. The customer or user
of the device
ZWave
Med
should ensure that it is used in such environment.
Immunity
Tests
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidelines
Voltage dips
in
accordance
with IEC
61000-4-11
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
The supply voltage quality must
correspond to that of a typical
commercial or hospital environment. If
the user of the device
ZWave
Med
requires continued operation, even in
the case of interruptions in the power
supply, it is recommended that the
device
ZWave
Med
be powered from an
uninterrupted power supply or a battery.
0 % U
T;
1 cycle and 70%
U
T;
25/30 cycles
Single phase: at 0°
0 % U
T;
1 cycle and 70% U
T;
25/30 cycles
Single phase: at 0°
Voltage
interruptions
accordance
with IEC
61000-4-11
0% U
T;
250/300 cycle
0% U
T;
250/300 cycle
Magnetic
field of
supply
frequency
(50/60 Hz) in
accordance
with IEC
61000-4-8
30 A/m
50 Hz oder 60 Hz
30 A/m
50 Hz
Magnetic fields at mains frequency
should have the typical values found in
a business or hospital environment.
Note: U
T
is the mains AC Voltage before application of the test level