Error Messages
Troubleshooting
Disposal
17
Page 31
Applicator not found
In the status bar, the message “
Handpiece not
found” appears.
Possible Cause
Handpiece not connected or incorrectly connected.
Fix for cause
Make sure the handpiece is properly connected. The plug must be properly snapped
in.
Device Malfunction
No response to the main switch / display remains dark
Possible cause 1
Mains power connection
Fix for cause 2
Check whether the power cord is properly plugged into the electrical outlet and
whether the plug is firmly inserted into the socket of the device.
Check the power cord for damage.
Check the power supply and mains socket.
Possible cause 2
Fuse
Fix for cause 2
In the power input socket of the device, there are micro-fuses that disconnect the
power supply in the event of an electrical problem. Open the cover and check the
fuses. If necessary, replace the faulty fuse.
Replace the fuse only with fuses of exact the same name or equivalent.
First, check the power supply completely for possible errors.
If the fault persists, contact the service representative / customer service immediately.