Safety and Maintenance
14.2 Maintenance
14
Page 27
Before cleaning and maintenance, the device must always be switched off with the
main switch, and the power cord must be disconnected.
Checking the impact
dome
During the course of normal use, a deformation/shortening of the rear impact dome
on the applicator head may occur. On a small scale, this will have no effect on
functionality.
In cases of more extensive deformation of the dome or considerable shortening, the
applicator head must be replaced.
To check the condition of the applicator head, a test template is supplied with the
device, which can be used to check whether the wear limit has been reached.
Depending on the frequency of use, the device should be inspected at least once per
month.
To do so, the applicator head is removed and measured with the enclosed test
template.
Air gap
➔
Applicator OK
Test template
Test template
Template is lying or air gap at the top
➔
Wear limit reached