System set-up
6.1 Mounting the cables
6
Page 9
Note:
Note:
Note
Silicone sleeve:
Note:
One active and one neutral electrode (common electrode) are needed for the
therapy.
Independent of which method is selected for performing the treatment, power
is always emitted via the ACTIVE output and the patient electrical circuit is
closed via the neutral electrode.
The use of the silicone sleeve allows the handpiece to be guided more
conveniently by the user. This is pulled with the wider side onto the electrode
connection on the handpiece.
ThermoTK offers various methods for performing the treatment.
Dynamic1:
1 dynamic electrode (only use “active” port!, otherwise the
dynamic electrode will not be detected)
+ 1 static common electrode
Dynamic2:
2 dynamic electrodes
Hands-free: 2 static self-adhesive electrodes
Ensure that the device is placed on a stable surface.
Note:
Make sure that the power switch of the device is set to “0”.
Connect power cable Connect the power cable to the provided port (13) on the device and connect
the cable to the mains.
Note:
The device may only be connected to power outlets with a protective contact.
Dynamic 1 method
1 dynamic ele 1 static common electrode
Connect handpiece Connect the handpiece with the white connection to the white ACTIVE
connection port.
Note:
Ensure that an electrode is inserted in the handpiece and that it is positioned
correctly and completely on the handpiece.
Connect common
electrode with cable
Attach the plug connector of the cable to the tab of the common electrode
(reusable electrode or self-adhesive single-use electrode).
Connect common
electrode cable
Connect the common electrode cable with the grey connection to the grey
NE connection port.