12.2. Vaco device
Vaco serial number:………………………………….
Serial number of board:……………………………..
Software version:…………………………………….
Visual inspection:
2x Fuses (0,4AT fuse at 100-120V device; 0,25AT fuse 230-240V device)
O
ok
100-120V device, jumpers JP1001, JP1004 are set.
O
ok
230-240V device, jumper JP1002, JP1003 are set
O
ok
Power switch and cable
O
ok
Internal wiring
O
ok
All screws present and tight
O
ok
Protective conductor present and tight
O
ok
Housing (no damage, labels legible, no dirt)
O
ok
Electrical safety:
Ground leakage current (max.500
µ
A)
O
ok
Ground leakage current 1st error (max.1000
µ
A)
O
ok
Housing leakage current (max.100
µ
A)
O
ok
Housing leakage current 1st error (max.500
µ
A)
O
ok
Patient leakage current (channel 1 max.100
µ
A)
O
ok
Patient leakage current 1st error (channel 1 max.500
µ
A)
O
ok
Patient leakage current (channel 2 max.100
µ
A)
O
ok
Patient leakage current 1st error (channel 2 max.500
µ
A)
O
ok
Function test:
Communication Vaco with Soleo
line
O
ok
Device test:
Stimulation current channel 1
O
ok
Stimulation current channel 2
O
ok
Stimulation current channel1 with suction electrodes (GA01,CV)
O
ok
Stimulation current channel2 with suction electrodes
(GA01,CV)
O
ok
Vacuum test/ tightness test:
(close all device openings, therapy vacuum >40kPa, pump start)
No repeated pressure regulation
O
ok
Pulsation
O
ok
Test run:
O
ok
All items checked
O
ok
Date:
______________________________
Name:
_______________________________
Signature:
____________________________
Please observe the legal regulations for repairs and maintenance of medical devices in your country.