Soleo
line
Service
Rev. B
26.07.10
-21-
K:\GRUPPE\QS\MDD-Doku\Service Handbuch soleoline.doc
8.3.2.2. Replacement of the pump
Unplug power cable from power source.
Unplug all cables from the housing.
Opening the cover.
Unplug the pump from the main board.
Detach the silicone tube from the pump.
Now unscrew the 3x screws M4x 8 mm located at the bottom of the housing as shown in Figure 11.
Figure 11
8.3.2.3. Replacement of the water separator
Unplug power cable from power source.
Unplug all cables from the housing.
Opening the cover.
Unplug the valve, water level sensor and the stimulating current cable from the main board.
Unplug the silicone tubes.
Now unscrew the 6x PT (Ejot) screws 3,0x12 mm located at the bottom of housing as shown in Figure 12