Service
Rev. B
26.07.10
-18-
K:\GRUPPE\QS\MDD-Doku\Service Handbuch soleoline.doc
8.2.
Accessory List
The valid accessory list is in the User manual implemented.
If accessories are combined, note the approved combinations and the relevant legal regulations. If third-party
accessories are used, this will result in a new medical system and the user will be solely responsible for the correct
combination or modification.
8.3.
Replacement of the C
omponents
Safety Note.
Do not touch modules or parts in the device when working on an open device with the power cable plugged in.
Electrical cables and wires may have a voltage up to 120V/ 240 V. Unplug power cable before opening and
closing the device
.
8.3.1. Replacement of Soleo
line
components
8.3.1.1. Opening the cover
! Caution mains voltage inside
1. Unplug power cable from power source.
2. Unplug all cables from the housing.
3. Removing the Cover
Unscrew the 7 Screws (PT EJOT 4,0x 14mm) located at the bottom of the housing.
Remove the cover carefully and fold it to the rear
Unplug the speaker cable and the power supply connectors from the main board.
Now housing can be placed aside.
8.3.1.2. Mount the cover
Before closing the housing always check the following:
The board and all other components are fastened in place.
All connectors are fixed.
No loose parts (such as screws) in the housing.
Mount the cover
Plug all cables to the main board
Place the cover on the housing and fix the cover with 7 Screws (PT EJOT 4,0x 14mm) located at the
bottom of the housing.