A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
4
Limb Occlusion Pressure (LOP) Button:
Controls the LOP feature. When the pulse sensor is in place
and cuff applied, pressing this button will start the process to
measure the patient’s LOP and give the user a recommended
tourniquet pressure (RTP).
AC Indicator Light Icon (Green LED):
Indicates unit is operating on AC Mains. This is the normal
means of operation (battery power is only intended for
emergency power loss or patient transport).
Battery Indicator Light Icon (Orange LED):
Indicates unit is operating on backup battery. This indicator
always flashes.
Alarm Indicator Light Icon (Red LED):
Visual indicator to show the unit is in an alarm condition.
LOP Heart Indicator Icon (Yellow LED):
Visual indicator to show the LOP function has been invoked.
Cuff Pressure Range:
50–475 mmHg, 1 mmHg increments
Pressure Accuracy:
±3 mmHg (50–475 mmHg)
Pressure Regulation:
±4 mmHg of set point
(10 second average under non-transient conditions without
external leaks)
Maximum Pressure:
475 mmHg cuffs
Time Alarm Set Range:
5–240 minutes; 1 minute increments
Timer Accuracy:
0.25% of elapsed time
Internal Diagnostics:
Program, memory, watchdog timer, transducer calibration,
improper valve actuation.
SIZE:
Height:
13.0 in. (33.02 cm)
Width:
9.5 in. (24.1 cm)
Depth:
10.375 in. (26.35 cm) (including clamp and ports)
Weight:
16.3 lbs. (7.4 kg)
DISPLAYS:
The
A.T.S.
3000 uses a backlit 1/4 panel LCD.
Pressure Display: Displays pressure setting, sensed cuff
pressure, and hardware failure conditions / other
messages.
Time Display: Displays time alarm set point, elapsed
time, and hardware failure conditions / other messages.
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
LOP