A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Table 3.2 Troubleshooting
SYMPTOM
POSSIBLE CAUSES
7. No cuff pressure reading.
a) Transducer amplifier not working.
b) Internal tubing kinked.
c) Transducer tubing on incorrect transducer.
8. Pump will not stop running.
a) Leak in internal hose or connector.
b) Internal tubing kinked.
c) Transducer(s) not working.
d) Transducer tubing on incorrect transducer.
9. Battery Fail alarm/message.
a) Blown battery fuse (board mounted F1).
b) Broken battery wire harness.
c) Dead or depleted battery.
10. Backup battery not charging.
a) Blown battery fuse (board mounted F1).
b) Battery not properly plugged into P10.
c) Unit not plugged into wall outlet (verify the
green AC indicator is illuminated).
d) Mains AC harness not properly plugged into P2
(verify the green AC indicator is illuminated).
e) Unit was not permitted to charge for at least
24 hours.
f) Defective battery.
g) Defective battery charging circuitry.
11. AMP FAIL alarm.
a) Transducer(s) amplifier out of range.
b) Battery fully depleted or defective.
c) Extremely high pressure exerted on transducers.
12. Unit cannot be set to STANDBY.
a) ON/STANDBY membrane not properly plugged
into P11.
b) Pressure sensed in the Main or Second Cuff
(unit will alarm “CUFF” “NOT” “DEFL”).
c) ON/STANDBY not fully pressed.
d) ON/STANDBY button defective.
13. Unit does not turn ON.
a) Membrane Panel not properly plugged into P8.
b) ON/STANDBY button defective.
c) Blown fuse(s).
d) Unit not plugged in and battery fully depleted.