A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
For 0.5 cycle
40% UT
(60% dip in UT)
70% UT
(30% dip in UT)
<5% UT
(>95% dip in UT)
for 5 s
<5% UT
(>95% dip in UT)
For 0.5 cycle
(120 Vac 60 Hz / 230 Vac
50 Hz)
40% UT
(60% dip in UT)
(120 Vac 60 Hz / 230 Vac
50 Hz)
70% UT
(30% dip in UT)
(120 Vac 60 Hz / 230 Vac
50 Hz)
<5% UT
(>95% dip in UT)
for 5 s
(120 Vac 60 Hz / 230 Vac
50 Hz)
Main power quality should be that of a typical
commercial or hospital environment. If the user of
the
A.T.S.
3000 requires continued operation during
power mains interruptions, it is recommended that
the
A.T.S.
3000 be powered from an uninterruptible
power supply or a battery.
The
A.T.S.
3000 is equipped with a backup battery
in the event of power mains interruptions. Consult
the Operator and Service Manual for additional
specifications.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Table 2.5 EMC Guidance and Declaration - EMC Immunity/Disturbances Continued