31
A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
31
NOTE: USE TP4 (DC COMMON) AS THE VOLTAGE REFERENCE FOR ALL MEASUREMENTS
LISTED ABOVE.
CAUTION: HIGH VOLTAGE ELECTRICAL HAZARD. HIGH VOLTAGE WILL BE
PRESENT ON THE POWER INPUT MODULE AND CONTROL BOARD. ALL SERVICE
WORK MUST BE COMPLETED BY QUALIFIED TECHNICIANS.
Note: If the unit is locked in a failure, not all test voltages will be valid. Above voltages are listed for normal
readings because failure readings will likely be unpredictable.
Table 3.3 Expected Test point Readings
Board Location
Nominal-Reading
Tolerance
Description / Comments
TP1
5.0 Vdc
0.0 Vdc
+0.2
5 Volt DC supply for battery charge circuit
0 Vdc while on battery operation
TP2
26.9 Vdc
+1.5
Battery DC bulk charging voltage
12 Vdc nominal while on battery operation
Voltage will vary while on battery operation and condition
of battery
TP3
N/A
N/A
Not used for any measurement
(high voltage DC common only)
TP4
DC Common
+50 mV
DC common supply
TP5
5.0 Vdc
+0.2
5 Volt DC switch mode supply
TP6
13.6 Vdc
+1.0
Nominal system voltage
Voltage will vary while on battery operation and condition
of battery
TP7
14.0 Vdc
+1.0
Bulk system voltage
0 Vdc while on battery operation
TP8
N/A
N/A
Development programmable pins only
TP9
N/A
N/A
Development programmable pins only
TP10
5.0 Vdc
+0.2
Nominal voltage under normal use
Voltage goes to 0 Vdc under system reset condition
TP11
4.1 Vdc
+0.2
ON / STANDBY switch logic
TP12
5.0 Vdc
+0.3
5 Volt DC regulator supply
TP13
Removed
Removed
Not used
TP14
5 Vdc
5 Vdc
Transducer reference voltage supply
TP15
Removed
Removed
Not used
TP16
13.6 Vdc
Nominal system voltage
Voltage will vary while on battery operation and condition
of battery
TP17
4.095 Vdc
+50 mV
Precision reference voltage supply