A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
17
Table 2.1 Alarm Condition
CONDITION
PRESSURE
DISPLAY
TIME
DISPLAY
APPROPRIATE ACTION/REMARKS
LINE OCCLUSION
An occlusion is present in the
cuff tubing
LINE OCCL
normal
Check for hose kinks or other defects.
LOW BATTERY VOLTAGE
Low battery voltage
normal
BATT LOW
PLUG IN
Plug unit in. If the unit is not plugged in, a battery
failure condition will occur and the unit will shut down
in a fail safe mode closing all valves.
While running
with a Low Battery Voltage Alarm Condition other
alarm conditions can not be guaranteed.
CHECK CUFF
An increase in cuff pressure has not
been sensed within 1 second after
depressing the inflate key
CHK
CUFF
Check to see that a cuff and hose assembly is
connected to the proper ports on the unit.
BATTERY FAILURE
Battery voltage is too low to ensure
proper operation
BATTERY FAIL
Plug unit in and cycle the ON/STANDBY Button.
CALIBRATION OUT OF SPEC
The transducer calibration is out
of specification
CAL M or CAL 2
FAIL
CAL M
(CALIBRATION MAIN)
or CAL 2
(CALIBRATION SECOND)
indicates a cuff
transducer circuitry is out of calibration. Pressure in
error by at least 6 mmHg will cause these failures.
Calibrate the unit.
CALIBRATION OUT OF SPEC
A transducer calibration is out
of specification
CAL FAIL
Indicates general calibration fail. Calibrate the unit.
AMPLIFIER FAILURE
Amplifier is out of range
AMP FAIL
Cycle the ON/STANDBY Button. If problem persists,
service the unit.
MATH FAILURE
Result of math operation was out
of range
MATH
FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
ROM FAILURE
Microprocessor failed a ROM
memory check
ROM
FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
RAM FAILURE
Microprocessor failed a RAM
memory check
RAM
FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
VALVE FAILURE
Improper valve combination
occurred
VALVE
FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
WATCHDOG FAILURE
Windowing watchdog system
detected a malfunction
WDT
FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
SYSTEM FAILURE
SYSTEM FAIL
Cycle the ON/STANDBY Button. If problem
persists, service the unit.
Continued