54 |
Persona The Personalized Knee
Surgical Technique
8 mm IM Step Drill
00-5978-014-00
Persona 8 mm IM Rod
42-5099-002-00
Persona Modular Handle
42-5099-014-00
Persona 5° Valgus
Alignment Guide
42-5099-005-05
Persona Fixed Resection Tower
42-5099-006-00
Persona 0° Distal Cut Guide
42-5099-010-00
Resect Distal Femur – Fixed Distal
Resection Instrument
(cont.)
Establish Femoral Alignment
Drill the IM canal using the 8 mm IM step drill
(Figure 111). Suction the canal to remove medullary
contents.
Insert the IM rod and assembled distal resection
instrumentation into the IM canal far enough to ensure
the most accurate replication of the anatomic axis.
Set the orientation of the fixed valgus guide by placing
it against the most prominent distal condyle and
rotating it about the IM rod so that it is aligned with
the epicondylar axis (Figure 112).
Appendix C: Optional Instruments
(cont.)
Figure 111
Figure 112
Figure 113
Note:
Setting rotation of the fixed valgus guide
is important for creating a distal resection that
matches the desired valgus angle selected. It does
not set the rotation of the femoral component.
To secure orientation, impact the captured pin on the
medial side of the fixed valgus guide until the head of
the captured pin is flush with the plate (Figure 113).