24 |
Persona The Personalized Knee
Surgical Technique
Persona Tibial Sizing Plate Handle
42-5399-017-00
Persona Cemented Tibial
Sizing Plate Size F Right
42-5399-075-02
Multi Pin Puller
00-5901-022-00
Alignment Rod with Coupler
00-5785-080-00
2.5 mm Male Hex Driver
42-5099-025-00
25 mm x 2.5 mm Female Hex Screw
42-5099-025-25
25 mm Shorthead Holding Pin
00-5977-056-03
Male-headed Screws/Pins must be removed from these
holes for Tibial Articular Surface Provisionals (TASP)
1. Depress
3. Release
2. Insert
Establish Size and Rotation of Tibia
Use only Persona tibial sizing, broaching, and provisional
instrumentation for preparation of Persona Implants.
Once tibial osteophytes have been thoroughly
removed, select the appropriate right or left sizing
plate that provides the desired tibial coverage, without
overhang at any location. Appropriate tibial sizing is
important as an over sized tibia component can result
in overhang, soft tissue impingement and pain, or
with stemmed components potential distal conflict
between stem and bone.
Attach the tibial sizing plate handle to the cemented
tibial sizing plate (Figure 49a). The recommended
tibial rotational alignment is within 5 degrees of the
axis created by the medial 1/3 of the tibial tubercle
and the PCL attachment point. The engraved lines
on the cemented tibial sizing plate can be used to
aid in establishing the desired tibial rotation. Rotate
the cemented tibial sizing plate to attain the desired
tibial rotational alignment. The notch in the lateral
periphery of the sizing plate is used to establish proper
position with respect to the lateral border of the tibia
without medialization of the sizing plate.
When the desired position has been attained, secure
the cemented tibial sizing plate by placing 25 mm x 3.2
mm (2.5 mm female hex) screws or 25 mm x 3.2 mm
short head holding pins in the medial and lateral holes
near the PCL cutout of the cemented tibial sizing plate
(Figure 49b). The remaining adjunct fixation holes shown
on the surface of the cemented tibial sizing plate can be
used if necessary. If the cemented tibial sizing plate is
to be used as a provisional in later steps, male-headed
screws/pins used in these holes must be removed prior
to using the Tibial Articular Surface Provisionals (TASPs)
(Figure 50). Ensure that the cemented tibial sizing plate
remains in the proper position when securing it to the
bone. Once desired alignment has been verified with the
alignment rod, remove the tibial sizing plate handle from
the cemented tibial sizing plate.
Figure 49b
Figure 50
Figure 49a