40 |
Persona The Personalized Knee
Surgical Technique
Persona TASP Right
EF +6 Bottom
42-5270-005-15
Persona TASP Top Right
Ultracongruent 4-11/EF
42-5272-005-10
Persona TASP Top Right
CR 3-11/EF
42-5270-005-10
Persona TASP Top Right
Medial Congruent 6-7/EF
42-5271-007-10
Persona Tibial Sizing
Plate Handle
42-5399-017-00
Persona TASP Shim EF
10 mm
42-5279-005-00
5-15 degrees
Tibial Articular Surface Provisional
(TASP) Assembly
(cont.)
The shims (10, 11, 12, 13, and 14 mm) are not side-
specific. Attach the tibial sizing plate handle to the
appropriate 10 mm shim (Figure 85a). While holding
the TASP top and bottom together with one hand,
lock the TASP top and bottom together by inserting
the appropriate 10 mm shim with the tibial sizing
plate handle (Figure 85b). The 10 mm shim will create
a TASP construct which matches the thickness of the
thinnest tibial bearing implant, 10 mm (Figures 86a
and 86b).
Technique Tip:
During assembly of the TASP
construct, slide the shim in using a direct anterior
approach between the TASP top and bottom.
To avoid inadvertent separation, maintain slight
pressure between the TASP top and bottom while
inserting the shim.
The shims are incremented by 1 mm to create TASP
constructs of 10 mm, 11 mm, 12 mm, 13 mm, or 14 mm
to match the implant offering. The +6 mm bottoms are
included for instances where the TASP construct needs
to be 16 mm, 18 mm, or 20 mm. In these circumstances,
the 10 mm, 12 mm, and 14 mm shims are to be used to
create the respective TASP constructs.
Note:
The maximum thickness of available CR
implants is 18 mm. MC, UC, and PS implants are
available in thicknesses up to 20 mm. Also, 15 mm,
17 mm, and 19 mm thicknesses are NOT available
in CR, MC, UC, or PS.
It is recommended that the thinnest TASP construct
(10 mm) be inserted into the joint space, with the knee
at greater than 30 degrees of flexion (Figure 87a), to
perform an initial ROM assessment.
If a thicker construct is needed to appropriately
fill and balance the joint space, place the knee in
approximately 5–15 degrees of flexion (Figure 87b)
to facilitate
in vivo
removal and insertion of the shims
with the tibial sizing plate handle.
Figure 85b
Figure 86b
Figure 86a
Figure 87a
Figure 87b
1. Depress
2. Insert
3. Release
> 30 degrees
Figure 85a