45 |
Persona The Personalized Knee
Surgical Technique
Persona UC Tibial Articular
Component Size 4-11/EF Right
42-5212-005-10
Persona PS Tibial Articular
Component Size 6-9/EF Right
42-5214-007-10
Persona Articular Surface Inserter
42-5299-001-00
Articular Surface Removal Instrument
00-5977-021-00
Persona Stemmed
Cemented Tibial
Component Size F Right
42-5320-075-02
Persona CR Tibial
Articular Component
Size 3-11/EF Right
42-5210-005-10
Persona MC Tibial
Articular Component
6-7/EF Right
42-5221-007-10
Implant Components
(cont.)
Bearing
(cont.)
Engage the hook on the articular surface inserter with
the mating slot in the front of the base plate and close
the lever with your index finger. This locks the inserter
to the tibial plate (Figure 96a). Squeeze the handle
of the articular surface inserter to seat the bearing
(Figure 96b). Open the lever and remove the articular
surface inserter.
Alternatively the bearing can be locked into the
tibial plate as described above, prior to tibial plate
implantation. If PS femoral implants are used, the
femoral implant should be implanted before the
tibial implant to avoid the excessive joint distraction
required to clear the post while implanting the PS
femoral implant. If CR femoral implants are used, the
tibial implant should be implanted prior to the femoral
implant to facilitate the removal of excess cement
from the posterior aspect of the tibia, prior to femoral
implantation.
Figure 96b
Figure 97
Figure 96a
Technique Tip:
Only
in vivo
assembly of the
ultracongruent bearing is recommended because
the design of the ultracongruent bearing precludes
cement removal from the PCL cutout area at the
tibial baseplate/bone interface.
Assemble the quick connect handle to the tibial
impactor head. Unlock collar and hold, insert quick
connect handle into tibial impactor head, release
collar, and rotate handle until “click” is heard. Position
the tibial plate bearing construct onto the tibia and
use the tibial impactor to impact it until fully seated
(Figure 97). Thoroughly remove any excess cement in
a consistent manner.
Technique Tip:
Do NOT impact or lever the articular
surface inserter or extractor tools when they are
attached to the tibial plate as this may disrupt the
fixation of the tibial plate to the bone and/or cause
damage to the implant or instrument. Also, do not
impact the bearing.
To remove a bearing from the baseplate, engage
the hook on the articular surface removal instrument
with the mating slot on the front of the baseplate and
squeeze the handle to disengage the bearing from the
base plate. The articular surface removal instrument
should not be used for provisional removal.