Section 7.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 93 of 220
WARNING
To protect the health of patients and staff against high radiation dosages, the
manual dose rate setting
mode remains blocked until you have initiated radiation
in one of the fluoroscopy modes with automatic exposure rate control at least once.
Only use the
manual exposure rate setting
mode in exceptional circumstances.
The automatic exposure rate control provides optimum image quality whilst
minimizing the dose rate.
To generate an image in the manual exposure rate setting mode, do the
following:
•
Select the desired organ program
• Select one of the fluoroscopy modes with automatic exposure rate
control:
•
Continuous Fluoroscopy
•
Pulsed Fluoroscopy
•
Snap Shot
•
Initiate radiation for a brief moment.
• Press
the
Manual Exposure Rate Setting
button on the mobile stand
and touch screen control keyboard. The kV value that has been
automatically determined is saved for subsequent fluoroscopic
initiations, and the system switches to the manual mode.
• Select the desired kV and mA values using the kV
Up Arrow
and
Down Arrow
buttons located on the Mobile stand control keyboard.
Each time you press an arrow button, the value is increased/decreased
by 1 kV, and the mA follows the preset kV/mA curve for the selected
anatomical program selected.
• Initiate
radiation.
To deactivate the manual exposure rate setting mode, do the following:
• Press
the
Manual Exposure
button on the mobile stand control
keyboard or on the DeskView touch screen.
Содержание QUANTUM
Страница 1: ...ZIEHM QUANTUM USER S MANUAL...
Страница 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Страница 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Страница 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Страница 51: ...Section 5 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 51 of 220 5 1 2 TRANSPORT POSITION FOR ZIEHM QUANTUM...
Страница 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Страница 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Страница 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Страница 165: ...Section 10 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 165 of 220 Fig 10 13 Configuration Screen...
Страница 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Страница 185: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 185 of 220 Fig 12 1 Focal Spot Reference...
Страница 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Страница 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...
Страница 204: ...Section 15 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 204 of 220 Fig 15 2 Generator Laser Alignment Labels...