8.16.2.
POSITIONING AID
The laser targeting device allows you to locate and position the C-arm to anatomical
locations on the patient without having to initiate radiation:
Press the
Laser
button. A laser-beam crosshair is generated, the central point of which
corresponds to the position of the central X-ray beam.
Position the C-arm in such a way above the patient that the central point of the laser-
beam crosshair is exactly in the center of the region to be screened.
8.17
DOSE DISPLAY FOR AKR AND ACCUMULATED AIR KERMA
8.17.1.
AKR /Cumulative Air Kerma Dose Display (Active only on system manufactured after June
10, 2006)
The
ZIEHM QUANTUM
calculates the dose values based on the kV, mA as well as time of the
radiation exposure. The display automatically changes between live display of
Air Kerma
(AKR)
or the
Cumulated Air Kerma
values (
AKR x Fluoroscopic Time
) automatically sets the
decimal point according to the dose rate and the accumulative values depending on weather
radiation is on or off. This also includes the dose for all images which have not been saved but are
part of the radiation accumulated since the last reset of the
Dose/Timer
display. See also
Section
12
technical information of this manual.
8.17.2.
DISPLAY FUNCTIONS:
The AKR/Cumulative air Kerma display performs two functions for displaying the dose the
patient is receiving or has received.
•
Radiation on: The system will calculate the
AKR (Air Kerma Rate)
or live dose rate based
on the displayed kV, mA. The air Kerma is displayed in units of
mGy/min
and will continue
to display at least once every two seconds until the user /operator releases the radiation
switch.
•
Radiation is turned off: The system will display the cumulated air Kerma dose value within 5
seconds with units of
mGy
.
8.17.3.
DOSE DISPLAY USER/OPERATOR
Using and resetting the Dose display:
Содержание QUANTUM
Страница 1: ...ZIEHM QUANTUM USER S MANUAL...
Страница 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Страница 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Страница 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Страница 51: ...Section 5 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 51 of 220 5 1 2 TRANSPORT POSITION FOR ZIEHM QUANTUM...
Страница 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Страница 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Страница 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Страница 165: ...Section 10 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 165 of 220 Fig 10 13 Configuration Screen...
Страница 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Страница 185: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 185 of 220 Fig 12 1 Focal Spot Reference...
Страница 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Страница 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...
Страница 204: ...Section 15 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 204 of 220 Fig 15 2 Generator Laser Alignment Labels...