Section 3.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 31 of 220
3.2.2.5.
FLAMMABLE GAS
Never use the system in the presence of flammable anesthetics, or other flammable gasses,
liquids, or vapors. Gasses and vapors can be ignited by electrical arcs that can occur during the
normal operation of power contacts, switches, circuit breakers, and other circuit components.
1.
Before the system is turned on
. If flammable liquids, gasses, or vapors
are present before
the system is turned on, do not unplug the system and do not turn it on!
2.
After the system is turned on
. If flammable substances are detected after the system has
been turned on, DO NOT turn the system off! DO NOT touch any of the controls, switches,
or knobs! DO NOT unplug the system! Remove all personnel immediately and then
ventilate the room. Remove any liquids that are producing flammable vapors to a safe
storage area.
3.3
MECHANICAL AND ELECTRICAL WARNINGS
Operate moveable assemblies and parts with caution. Inspect moveable assemblies and parts at
the intervals stated in this manual.
To prevent un-wanted movement of the mechanical c-arm assembly, always maintain control by
holding on to and then slowly releasing the mechanical locks.
Transport C-arm system as outlined in this manual. Never transport the C-arm without first
locking the mechanical brakes and Monitor support arm into the transport position. Failure to
perform the necessary steps to position and then lock the C-arm for transport can result in
death
or serious injury
to the operator and or patients.
3.3.1
ELECTRICAL SAFETY
Only allow properly trained and qualified service personnel to access any internal parts. Live
electrical terminals are deadly! Be sure that line disconnect switches are open (off) and that
precautions are taken before opening access doors, removing enclosures and panels, and attaching
accessories. Do not remove access covers from the generator, control unit, or main power
assembly until the main and auxiliary power supplies have been disconnected. If you fail to
follow these instructions, you may be seriously injured or killed. There is always high voltage in
certain components of the system’s equipment,
even when the system is disconnected from a
power source. Touching these components can cause
death or serious injury.
Any Supplementary, third-party devices, and or components used in combination with the
ZIEHM QUANTUM
must comply with the safety requirements according to IEC 60601-1
and/or IEC 60601-1-1 or furnish proof of an equivalent degree of safety. Combining the
ZIEHM
QUANTUM
with equipment which does not comply with IEC 60601-1 and/or IEC 60601-1-1,
will lower the safety classification, compromising the safety of the whole system, which may
result in
serious injury or even death!
Non-compliance with IEC 60601-1 and or IEC 60601-1-1 may also result with invalidation of
warranty!
3.3.2
GROUNDING
The equipment may only be connected to power supplies with separate equipment grounding.
3.3.3
HIGH VOLTAGE COMPONENTS
High Voltage system Components (50-60Hz)
•
High-voltage transformer and x-ray tube assemblies (110,000 volts)
•
Power transformers (220-230 VAC and 120 VAC)
Содержание QUANTUM
Страница 1: ...ZIEHM QUANTUM USER S MANUAL...
Страница 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Страница 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Страница 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Страница 51: ...Section 5 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 51 of 220 5 1 2 TRANSPORT POSITION FOR ZIEHM QUANTUM...
Страница 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Страница 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Страница 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Страница 165: ...Section 10 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 165 of 220 Fig 10 13 Configuration Screen...
Страница 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Страница 185: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 185 of 220 Fig 12 1 Focal Spot Reference...
Страница 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Страница 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...
Страница 204: ...Section 15 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 204 of 220 Fig 15 2 Generator Laser Alignment Labels...