Section 9.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 131 of 220
The
CINE
Button will be highlighted on the DeskView screen and the selected frame rate
will be display on the touch screen.
3.
Press and hold the hand or foot switch. The left monitor will display the message
“INJECT” in the lower center of the screen to indicate that the system has started capture of
images.
4.
Start the injection of contrast media.
5.
Continuing holding the exposure switch until the desired vascular anatomy has been
visualized.
6.
The acquired images will begin to play back in a continuous loop from the hard drive or
from memory, depending on frame rate selected and the system’s optional equipment.
9.7.1.
ACQUIRING A NEW CINE RUN
If scout fluoroscopy is necessary between runs, press the
CINE
button to exit and return to
normal fluoroscopy.
Repeat the Cine procedure above. To acquire another study /cine run
9.7.2.
IMAGE STEP
During CINE replay, image playback can be stopped and individual
images can be reviewed and or stored to the patient file.
1.
Press one of the
Contrast
buttons on the Mobile Stand Control Panel or the
Arrow
playback
buttons on the DeskView Control Panel. This places the system into a playback
mode that steps through images one at-a-time.
1)
The message “F/R OFF” will display in the upper right corner of the monitor,
indicating that the image loop has stopped and that the operator can now step
through the images.
Press either button to step through the images. Hold down either button to step
through the images at nearly the selected frame rate.
2)
When image playback has stopped, select the image to transfer to the patient file
and press either of the
Image Store
buttons.
3)
The message “Stored to XXXX” is displayed on the left monitor which indicates
that the image has been copied to the patient file as image number XXXX.
Содержание QUANTUM
Страница 1: ...ZIEHM QUANTUM USER S MANUAL...
Страница 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Страница 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Страница 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Страница 51: ...Section 5 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 51 of 220 5 1 2 TRANSPORT POSITION FOR ZIEHM QUANTUM...
Страница 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Страница 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Страница 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Страница 165: ...Section 10 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 165 of 220 Fig 10 13 Configuration Screen...
Страница 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Страница 185: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 185 of 220 Fig 12 1 Focal Spot Reference...
Страница 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Страница 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...
Страница 204: ...Section 15 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 204 of 220 Fig 15 2 Generator Laser Alignment Labels...