Section 4.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual
Page 45 of 220
5.
75 kV
tube voltage is set;
6.
Radiographic time is set to One second.
The control panel keys on the Mobile stand, shown in Figure 4.2, will light up.
Keys Activated in Default Mode.
Mobile stand keyboard
DeskView touch panel control
After 40 to 60 seconds of filament preheating (approximately), you may begin fluoroscopy by
pressing either the hand switch or foot switch. The kV/mA will be automatically adjusted and the
technique will be displayed on the C-Arm control panel display.
4.6.4
PERFORM THE FOLLOWING OPERATIONAL TESTS & VERIFY THAT THE
FOLLOWING RESULTS OCCUR:
1.
During fluoroscopy, the live image appears on the left monitor.
2.
When exposure occurs (either during fluoroscopy or radiography) the amber light on top of
the left monitor lights up, and the radiation indicator on the C-Arm control panel lights up.
3.
Open and close the slot and iris collimators. (Perform Collimation test with Fluoroscopy
on)
4.
After fluoroscopy is completed, the last image is held.
5.
Rotate the slot collimator during radiation to make sure it is operational.
6.
Reverse the image.
7.
Rotate the monitor image using the Image rotation buttons on the touch screen control of
the DESK VIEW.
8.
Select manual kV mode. Press hand switch to start radiation, while radiation is on press the
kV up key, the kV value must change.
WARNING:
To prevent the camera from being over-driven by high light-levels, do not perform
fluoroscopy in the kV manual mode at a kV levels higher than needed.
Содержание QUANTUM
Страница 1: ...ZIEHM QUANTUM USER S MANUAL...
Страница 2: ...Table of Contents MAN 06 0017H ZIEHM QUANTUM User s Manual Page 2 of 220...
Страница 10: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 10 of 220...
Страница 12: ...About this Manual MAN 06 0017H ZIEHM QUANTUM User s Manual Page 12 of 220...
Страница 51: ...Section 5 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 51 of 220 5 1 2 TRANSPORT POSITION FOR ZIEHM QUANTUM...
Страница 71: ...Section 6 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 71 of 220...
Страница 96: ...Section 7 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 96 of 220...
Страница 132: ...Section 9 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 132 of 220...
Страница 165: ...Section 10 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 165 of 220 Fig 10 13 Configuration Screen...
Страница 180: ...Section 11 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 180 of 220...
Страница 185: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 185 of 220 Fig 12 1 Focal Spot Reference...
Страница 196: ...Section 12 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 196 of 220...
Страница 202: ...Section 14 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 202 of 220...
Страница 204: ...Section 15 0 MAN 06 0017H ZIEHM QUANTUM User s Manual Page 204 of 220 Fig 15 2 Generator Laser Alignment Labels...