Guidance and manufacturer’s declaration – electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test speci
fi
ca-
tion for ENCLO-
SURE PORT
IMMUNITY to
RF wireless
communications
equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Modulation
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 – 390
TETRA 400
Pulse modulation
18 Hz
1.8
0.3
27
450
380 – 390
GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
704 – 787
LTE Band 13,
17
Pulse modulation
217 Hz
0.2
0.3
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
Pulse modulation
217 Hz
2
0.3
28
870
930
1720
1700 – 1990
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
217 Hz
2
0.3
28
1845
1970
2450
2 400 – 2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217 Hz
2
0.3
28
5240
5 100 – 5 800
WLAN 802.11
a/n
Pulse modulation
217 Hz
0.2
0.3
9
5500
5785
All rights reserved.
02392_EN_180x140_GM_V1.indd 22
02392_EN_180x140_GM_V1.indd 22
19.11.2019 15:54:19