• Do not use in the vicinity of high-frequency surgical equipment which is switched on and RF-shielded
rooms of an ME system for magnetic resonance imaging as the intensity of the electromagnetic interfer-
ence is high there.
• Mobile RF communications devices (including peripherals such as antenna cables or external antennae)
should only be used with a minimum gap of 30 cm from each part of this device, including the power
cable. Otherwise the power of this device may decrease.
• If the device does not work in the ultrasound functional test (as speci
fi
ed in the “Read Carefully Before First Use”
chapter) as a result of electromagnetic interference, it must not be used and an attempt should be made to use
it in a di
ff
erent place.
• To maintain the basic safety and performance capacity with regard to electromagnetic interference for the
expected lifespan, please take precautions if the place of use is in the vicinity of (i.e. less than 1.5 km from) AM,
FM radio or television aerials.
The appliance is complied with the following emission and immunity standard, and test level:
Emissions Test Standard
Compliance
RF emissions
CISPR 11
Group 1, Class B
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage
fl
uctuations /
fl
icker emissions
IEC 61000-3-3
Not Applicable