Service Instructions
Version 07, 2015-12-03
7 / 27
Order No. -
5
OPMI
pico
5.1
Removing, installing and adjusting the OPMI® pico illumination
module
Removal:
1)
Disconnect the light guide and (if necessary) the
system cable (video cable) from the OPMI®
microscope.
2)
Remove four screws (
1
) and cover (
2
).
Fig. 1: Remove cover
3)
Unscrew four screws (
3
) and remove OPMI®
pico illumination module (
4
).
Installation:
1)
Insert OPMI® pico illumination module (
4
) and
mount it by tightening four screws (
3
) only
slightly.
Fig. 2: Remove OPMI® pico illumination module
5.2
Exchanging a filter
1)
Swing out the filter to be exchanged (
5
) by
turning filter selection ring (
6
).
2)
Remove filter (
5
) from the mount and insert the
new filter.
Note
Make sure that the filter is absolutely clean:
Any contaminations present on the filter
are imaged in the illuminated field.
3)
Proceed in the reverse order for assembly.
Subsequent adjustment is required.
Fig. 3: Exchange the filter
2
1
4
3
3
6